ECG
Connectors: 3- or 5-lead
Leads: User-selectable; I, II, III, aVR, aVL, aVF, V (5-lead); I, II,
or III (3-lead)
Gain Selection: x1/2, x1, x2, and x4
Electrosurgery Protection: Yes
Defibrillator Protection: Yes
Pacer Detection/Rejection: Yes
Non-Invasive Blood Pressure
Technique: Oscillometric measure upon inflation
Automatic Measurement Cycles: 2, 3, 5, 10, 15, 30 min; 1, 2, 4 hr
Measurement Range: Adult: 30 to 300mmHg; Neonate: 0 to 150 mmHg
Resolution: 1mmHg
Transducer Accuracy: ±2mmHg or ±2%,
which ever is greater
STAT mode: 5 min of consecutive readings
Pulse Oximetry
Range: 1% to 99%
Resolution: 1%
Accuracy: 70% to 99% range: ±2%; 50% to 69% range:
±3%; <50%: unspecified
Method: Dual wavelength LED
Heart (Pulse) Rate
Source: ECG (primary); Pleth; NIBP
Range: 20 to 300 bpm (30-240 bpm, NIBP)
Accuracy: ±1bpm or ±1% (ECG); ±2 bpm
or ±2% (NIBP)
Temperature
Channels: 1
Range, Accuracy: 68° to 113° F (20° to
45° C): ±0.1° C
Probe: YSI-400 and YSI-700 series
Respiration
Rate: 6 to 150 breaths/min
Resolution: 1 breath/min
Accuracy: ±1% or ±1 breath/min, whichever is
greater
Source: ECG-TTI

Alarms
Indication: Audible; visual
Silence: Yes; 2 min or permanent
Trend Memory
Memory: 24 hr at 30 sec intervals
Display: Tabular, Graphical
Display
Screen: 10.4" active color TFT
Resolution: 640 x 480 pixels
Waveforms: 6, maximum
Languages: English, French, German, Italian, Portuguese, Spanish,
Russian, Chinese
Controls
Keys: 9; membrane-activated
Rotary Knob: Push and rotate; 24 steps/turn
System Outputs
Com Ports: RS 232-compatible; DB9 (Com 1)
Nurse Call: Contact switch; 3.5mm stereo phone jack, standard 14mm
length
Defib/Sync: BNC connector
Video Port: High density D-sub 15-pin
Printer
Type: Thermal line printer
Paper Speed: 12.5, 25, 50 mm/sec
Mechanical
Weight: 14 lb; 6.4kg
Size: Base unit: 11" (H) x 13" (W) x 10.3" (D); 28cm (H) x 33cm (W)
x 26.2cm (D)
Power Requirements
Voltage: 100, 120, 220, 240 VAC; 50/60 Hz
Consumption: 35 W, typical
Batteries: 1; sealed lead acid; life: 3 hr, typical; recharge time: 4.5
hr
Environmental
Operating Temperature: 59° to 95° F; 15° to
35° C
Storage Temperature: 23° to 122° F; -5° to
50° C
Operating Humidity: 15% to 90%, noncondensing
Classification
Medical Device: Class II Equipment (IIb EU)
Electrical Protection: Class I Equipment
Degree of Protection: Type CF, Defib-proof
Protection Against Ingress: IPX1
Specifications subject to change without notice.

|