ECG
Number
of Leads: 3
Lead Selection: I, II, III
Gain Selection: 0.5 mV, 1 mV, 2 mV, 4 mV, 8 mV full scale
Frequency Response: Diagnostic: 0.05 to 100 Hz Monitor: 0.5 to 40 Hz
Isolation: Conforms to IEC 601-1
Sweep Speed: 25, 50 mm/sec
Electrosurgery Protection: Yes
Defibrillation Protection: Yes
Pacer Rejection: Yes
Heart Rate Display and Alarms
Range: 30 to 300 BPM (ECG, SpO2 only)
Accuracy: > of ±1 BPM or 1%
Alarm Indication: Audible, Visual
Alarm Volume: Adjustable: 10 settings + Off, user-selectable
Alarm Silence: Yes, 2 minutes
Noninvasive Blood Pressure
Technique: Automatic oscillometric upon inflation
Max. Inflation Pressure: Adult 300mmHg; Neonate 150mmHg
Automatic Measurement Cycles: 1, 2, 3, 5, 10, 15, 30, and 60 min; STAT
pressure interval is 5 minutes
Impedance Respiration
Range: 9 to 150 breaths/minute
Resolution: 1 breath/minute
Accuracy: > of ±1% or ±1 breath/minute
Temperature
Range: 20 to 45° C, 68 to 113° F
Accuracy: ± 0.1° C
Display Resolution: 0.1° C or 0.1° F
Probe: YSI-400 Series
Oxygen Saturation (SpO2)
Averaging Time: Selectable 3 to 21 seconds
Search Time: Selectable, 10 to 40 seconds
Waveform: User Selectable
Gain: User Selectable
Sensor Type: Reusables: DOX™, MultiSite™,
SHELL™; Disposables: Adult, Pediatric, Infant, Neonatal
Audio: Beep with each detected pulse, pitch varies with oxygen
saturation
Range: 1 to 99%
Resolution: 1%
Accuracy: ±2% (70 to 99%); ±3% (50 to 70%)
Method: Dual wavelength LED
External Power Supply
Input Voltage: 90-260 VAC; 47-63 Hz
Output Voltage: 16V DC
Operating Environment
Temperature: 15° to 40° C (59° to 104°
F)
Humidity: 15 to 90% (non-condensing)
Storage Temperature: -5° to 50° C (23 to 122°
F)
Display
Main Display: Active color LCD or monochrome
Size: 4.76" (121mm) W x 3.58" (91mm) H
Resolution: 320 pixels W x 240 pixels H
Numerical Display: LED (green: SpO2%; red: Rate, Sys, Dia, Map)
Number of Traces: 3 maximum
System Outputs
Digital: DB-9 (serial interface)
Analog Output: BNC, ±1V range
Trends
Memory: 24 hours of tabular patient data at 3-minute NIBP interval and
1 minute other parameters
Battery
Type: Rechargeable lead acid
Life: 1 hour
Recharge Time: 6 hours to 100% capacity
Physical Properties
Size: 19 cm (7.6”) H x 26 cm (10.125”) W x 25 cm
(9.75”) D
Weight: 3.83 kg (8.5 lbs)
Classifications
Type of Protection: Class I equipment
Degree of Protection: Type CF defibrillator-proof equipment
Protection Against Liquids: IPX1, drip-proof equipment
Approved by UL, CSA,
CE.
Specifications subject to change without notice.
|